● 8 fps frame rate, high positioning accuracy
● Dynamic imaging
● Ingress Protection rating of IP65
●16MV high-energy protection design
● Direct evaporation coupled Cesium Iodide (CsI) Scintillator effectively reduces light scattering between pixels
X-Panel 3030a FQM-H is a 30 x 30cm fixed type, amorphous silicon, dynamic flat panel detector. It features excellent sensitivity and high dynamic range modes, the structural epitaxial radiation shield design offers user flexibility, and external protection. The 16MV high-energy is an optimal solution in tumor localization and radiotherapy accelerator application.
Technology |
|
Sensor |
a-Si |
Scintillator |
CsI / GOS |
Active Area |
286 x 286 mm |
Pixel Matrix |
2048 x 2048 |
Pixel Pitch |
140 μm |
AD Conversion |
16 bits |
Interface |
|
Data Interface |
Gigabit Ethernet |
Exposure Control |
Pulse Sync In / Pulse Sync Out |
Frame Speed |
8fps (1x1) / 16fps (2x2) |
Operating System |
Windows7 / Windows10 OS 32 bits or 64 bits |
Technical Performance |
|
Resolution |
3.5 lp/mm |
Energy Range |
≤16 MV |
Lag |
0.8% @1st frame |
Dynamic Range |
≥86 dB |
Sensitivity |
620 lsb/uGy |
SNR |
49 dB @(20000lsb) |
MTF |
72% @(1 lp/mm) |
44% @(2 lp/mm) |
|
25% @(3 lp/mm) |
|
DQE |
66% @(0 lp/mm) |
42% @(1 lp/mm) |
|
28% @(2 lp/mm) |
|
Mechanical |
|
Dimension(H x W x D) |
418 x 372 x 31.4 mm |
Weight |
9.1 Kg |
Sensor Protection Material |
Carbon Fiber |
Housing Material |
Aluminum Alloy |
Environmental |
|
Temperature Range |
10~35℃(operating);-10~50℃(storage) |
Humidity |
30~70% RH(non-condensing) |
Vibration |
IEC 60068-2-6 Operation: 0.7G, 10-150Hz |
Shock |
IEC 60068-2-27 Operation: 2G, duration 11ms |
Dust and Water Resistant |
IP65 |
Power |
|
Supply |
100~240 VAC |
Frequency |
50/60 Hz |
Consumption |
13W |